Default Settings and Biomedical Support
VF4 Infinity Gamma Series Page C-3
Adult
Settings
Pediatric
Settings
Neonatal
Settings
After patient
discharge,
settings
return to...?
ECG
Alarm OnOnOn Saved Setup
Alarm Limits
(upper/lower)
120/45 150/50 180/80 Saved Setup
Alarm Recording Off Off Off Previous
Patient
Tone Source ECG ECG ECG Previous
Patient
Tone Volume Medium Medium Medium Previous
Patient
Pacer Detect Off Off N/A Saved Setup
QRS Marks Off Off Off Factory
ECG Processing ECG 1&2 ECG 1&2 ECG 1 Previous
Patient
ECG Leads 5-Lead 5-Lead 3-Lead Previous
Patient
Arrhythmia*
ASY, VF, VT,
RUN, AIVR, SVT,
CPT, BGM Alarm
BRDY, TACH,
PAUS, ARTF
Alarm
On
Off
On
Off
n/a
n/a
Saved Setup
ASY and VF
Alarm Recording
Other Events
Record
Off
Record
Off
n/a
n/a
Saved Setup
Arrhythmia
Monitoring
Basic Basic n/a Saved Setup
*Not applicable for neonatal monitoring.

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
ManufacturerDräger
Device TypePatient Monitor
Weight5 kg to 10 kg (depending on the model and configuration)
ConnectivityEthernet, USB, WLAN (optional)
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
Alarm SystemVisual and audible alarms, customizable alarm limits
Device CategoryMedical Device
SeriesInfinity
DisplayColor TFT
Display Size12.1 inches
Power SupplyAC Power, Internal Battery
Battery LifeUp to 4 hours (depending on configuration)

Related product manuals