Network Applications
VF4 Infinity Gamma Series Page 3-15
Central Display
The central station identifies wireless monitors by a trans-
mission icon on the central display. Upon undocking, the
bed’s central display channel (viewport) remains assigned to the
wireless monitor. If another monitor docks on the same Docking
Station which is still associated with the wireless monitor, the
new monitor receives the same bed label from the Docking Sta-
tion, but does not replace the wireless monitor in the viewport of
the central station. In order to view the new monitor on the cen-
tral display, the user has to assign a different display channel to
the new monitor via the central station’s Assign Bed menu.
NOTES:
! Patient name and ID continuously identify the patient on cen-
tral displays, recordings, and data bases, whether the monitor is
in transport or docked.
! If a wireless monitor is transmitting to a central station but has
not been assigned a viewport at that station, the message Not
monitored by central appears in the monitor’s message area.
Patient Relocation
You can move monitor and patient to a different room and care
unit by simply docking the monitor at the new location. Alterna-
tively you can select care unit and bed label of the intended new
location via the Patient Admit menu (see the section Selecting
Care Unit and Bed Label).
During transport, the wireless monitor continues to communicate
with its assigned central station via the access points. Upon dock-
ing at the new location, the monitor accepts the new Docking Sta-
tion’s care unit and bed label assignments as well as its display
position on the central station’s cluster screen.
If you change central stations, that is, move the monitor to a
room that is monitored by a different central station, the original
central station blanks the monitors data and displays the message
Bed Disconnected.
NOTE: During the transition between wireless and wired network
operation, the network may generate an error tone, if the transition
takes more than a few seconds. There is no loss of data and normal
network operation continues, as soon as the M
ULTIVIEW WORK-
S
TATION has recognized the monitors new connection status.

Table of Contents

Other manuals for Dräger Infinity Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Infinity Series and is the answer not in the manual?

Dräger Infinity Series Specifications

General IconGeneral
ManufacturerDräger
Device TypePatient Monitor
Weight5 kg to 10 kg (depending on the model and configuration)
ConnectivityEthernet, USB, WLAN (optional)
Parameters MonitoredECG, SpO2, NIBP, Temperature, Respiration
Alarm SystemVisual and audible alarms, customizable alarm limits
Device CategoryMedical Device
SeriesInfinity
DisplayColor TFT
Display Size12.1 inches
Power SupplyAC Power, Internal Battery
Battery LifeUp to 4 hours (depending on configuration)

Related product manuals