en | Maintenance
64 X-am
®
2800 / X-am
®
5800
WARNING
Risk of explosion
To reduce the risk of ignition of a flammable or potentially explosive atmosphere,
observe the following:
Do not charge or replace the battery in potentially explosive atmospheres or
in explosion-hazard areas.
Only use the battery charger specified by Dräger. The use of a different
charger nullifies the explosion protection certification of the gas detector.
Insert the gas detector into the charger module.
If the gas detector is switched on, then it is automatically switched off after
being inserted.
5.8 Replacing the battery
WARNING
Risk of explosion
To reduce the risk of ignition of a flammable or potentially explosive atmosphere,
observe the following:
Do not open the gas detector in explosion-hazard areas.
Only the battery types specified in the technical data must be used.
Do not charge or replace the battery in potentially explosive atmospheres or
in explosion-hazard areas.
Check that the fastening screw of the battery pack is secure before use.
NOTICE
Damage to the gas detector
The power pack for the X-am 2500/5000/5100/5600 gas detectors (order
no. 8318704) must not be used for the X-am 2800, as the seal cannot ensure
the IP protection for the X-am 2800 and was not approved together with the X-
am 2800.
Only use the battery pack (order no. 3703887) with power pack 3703880.
1. Switch off the gas detector.
2. Undo the screw on the power pack.
3. Remove and replace the NiMH power pack T4 (type HBT 0010).
4. Insert the power pack into the gas detector and tighten the screw, the gas
detector switches on automatically.

Table of Contents

Other manuals for Dräger MQG 02 0 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger MQG 02 0 Series and is the answer not in the manual?

Dräger MQG 02 0 Series Specifications

General IconGeneral
BrandDräger
ModelMQG 02 0 Series
CategoryGas Detectors
LanguageEnglish

Related product manuals