5.9 Cleaning
The gas detector does not require any special care.
If it is heavily soiled, wash off the gas detector with cold water, using a sponge if
necessary. Dry the gas detector with a cloth.
NOTICE
Damage to the gas detector!
Harsh cleaning utensils (e.g. brushes), detergents and solvents can destroy the
dust and water filters.
► Only clean the gas detector with cold water and, where necessary, a sponge.
6 Device settings
Only trained and qualified personnel may change the device settings.
For more details, refer to the technical manual.
6.1 Factory settings
Different settings can be selected to meet customer requirements when
ordering. The setting can be checked and changed with the Dräger CC-Vision
PC software.
The changed parameter settings must be checked after being transferred to
the gas detector to ensure that the values have been transferred correctly.
Parameters that cannot be viewed on the gas detector must be read out and
checked after being changed using the Dräger CC-Vision PC software.
For information on suitable cleaning agents and disinfectants and their
specifications, see document 9100081 at
www.draeger.com/IFU.
Feature Setting
Fresh air calibration without password On
Bump test without password On
Operation signal (acoustic) Off
Bluetooth
®
Off
Switch off is allowed On
Capture range
1)
On