Device settings | en
X-am
®
2800 / X-am
®
5800 67
Sensor settings:
A1 alarm:
Latching
Acknowledgeable
On / Off
On / Off
A2 alarm:
Acknowledgeable On / Off
Alarm threshold A1 increasing
(in measurement module)
0 to A2
Alarm threshold A2 increasing
4)
(in measurement module)
A1 to full scale deflection
Alarm threshold A1 decreasing
(in measurement module, only O
2 sen-
sor)
A2 decreasing to A1 increasing
Alarm threshold A2 decreasing
(in measurement module, only O
2 sen-
sor)
0 to A1 decreasing
Interpretation type
1)
Inactive, TWA, STEL, TWA+STEL
Alarm threshold STEL (in measure-
ment module)
1)
0 – full scale deflection
Alarm threshold TWA (in measure-
ment module)
1)
0 – full scale deflection
1) Evaluation only if the sensor is designed for this.
2) Corresponds to averaging time and is used to calculate the STEL exposure value.
3) Corresponds to averaging time and is used to calculate the TWA exposure value.
4) Max. 60% LEL applies for ex-channels
Name: Area / setting