Assembly and preparation
34 Instructions for use Polaris 600
– To remove the handle, see the supplement to
the instructions for use concerning
reprocessing of Polaris.
Fitting the Dräger disposable handle
(optional)
Safety information concerning the Dräger
disposable handle
– The disposable handle is intended solely for
single use. The disposable handle must not be
reused, reprocessed, or sterilized.
– The membrane on the disposable handle must
not exhibit any cracks, holes, or other damage.
– The inner handle must be free from damage.
– To prevent soiling of the disposable handle, do
not unpack it until immediately before an
operation.
– Do not use the disposable handle if the
packaging is damaged.
– Do not use a knife to open the packaging for
the disposable handle.
Immediately before an operation, fit the disposable
handle under sterile conditions as follows:
1 Hold the disposable handle (A) by the recessed
grips (B) and fit it onto the inner handle.
WARNING
Risk of infection
Reuse of the disposable handle can lead to
cross-infection.
The disposable handle is intended solely for
single use. The disposable handle must not
be reused.
WARNING
Risk of personal injury and property damage
The handle can fall down into the operating
field if it is not fitted properly.
The handle may only be fitted by users.
36005
WARNING
Risk of infection
The disposable handle must be checked for
secure attachment after it has been fitted to
the inner handle.
The disposable handle must be free from
damage.
NOTE
To facilitate fitting the disposable handle, non-
sterile OR personnel may hold the light steady.