Instructions for use Polaris 600 81
Maintenance
Maintenance
Light system
The following table shows the preventive
maintenance intervals:
WARNING
Risk of faulty components
Device failure is possible due to wear or
material fatigue of the components.
To maintain the function of all components,
this device must be inspected and serviced at
the intervals specified by the manufacturer.
CAUTION
Risk of electric shock
All poles of the power supply must be
disconnected when carrying out service activities
on the light system. The switch required to do this
is a part of the mains connection component.
The mains connection component must be
prevented from being switched on again
unintentionally.
Component Interval Task Personnel responsible
Arm system Every 2 years Grease the brake
screws, see "Checking
the circlip on the lower
swivel arm to the spin-
dle joint" on page .
– Grease the brake
screws on the swivel
arm and the spring
arm.
Specialized service per-
sonnel