Maintenance
80 Instructions for use Polaris 600
Display mounts
Component Interval Task Personnel responsible
Display mount and arm
system
Every 2 years
Check that the segment
and the screws on the ceil-
ing tube/spindle connec-
tion are secure.
The connection must be
free from defects.
Specialized service per-
sonnel
Check that the cable clips
in the spring arm are
undamaged.
The cable clips must be in
the intended position.
Check that all 4 screws on
the socket / socket attach-
ment connection are
seated correctly.
Check for paint damage
and cracks in plastic parts.
Check whether all plastic
parts are present and in
the correct position.
Visually inspect the sup-
ply lines for pressure
marks or chafe marks.
Visually inspect the sup-
ply lines for abrasion wear
in the housing.
Check that the rating plate
is present and is readily
legible.
Perform functional and
visual inspection.