Instructions for use Polaris 600 71
Troubleshooting
Polaris 600 remote control for MedView
camera
Status LED (D)
Off Ambient light mode (Endo light)
is not switched on (OR light
mode is switched on)
Light is in standby mode
No remedial action nec-
essary
Continuously white
on the light control
panel
Continuously orange
on the wall-mounted
control panel
Light is in ambient light mode
(Endo light)
No remedial action nec-
essary
LED Status Cause Remedy
Fault Cause Remedy
Remote control is
not responding
(status LED lights
up green when
key is pressed)
Remote control is not within the reception range of the IR
receiver
Aim the remote control at
the IR receiver, see
"Operating the MedView
camera using the
Polaris 600 remote con-
trol for MedView camera"
on page 64
Hold the remote control
closer to the IR receiver,
see "Operating the Med-
View camera using the
Polaris 600 remote con-
trol for MedView camera"
on page 64
Room address of the remote control or of the MedView
camera is incorrectly set
Remote control and
MedView camera must
be set to the same room
address.
Notify DrägerService
Remote control is
not functioning
(status LED does
not light up when
key is pressed)
Battery is discharged Fit a new battery, see
"Replacing the battery in
the Polaris 600 remote
control for MedView
camera" on page 65
Battery is defective Fit a new battery, see
"Replacing the battery in
the Polaris 600 remote
control for MedView
camera" on page 65