1. Press the corresponding numeric key to select appropriate variables.
2. Select value you wish to enter, using the number key on the keyboard. Then
press Enter.
Note: The Date and Time data are non-volatile memory even power off. The memory
will lose if the on board battery is out of power.
3.8.3 Recording Lamp Usage
Press <3> on the keyboard from the Calibration submenu, time setting submenu
will appear on the right of the screen as shown below.
11/16/02 15:25 600.0 nm 0.854 ABS
Calibration
1 WL Calibration 1 Change New WI Lamp
2 Set Time 2 Change New D2 Lamp
3 Lamp Time WI Lamp Usage (h) 10
4 GLP/GMP D2 Lamp Usage (h) 10
Select Function: _
The Lamp Time setting keeps a record of how many hours of the D2 or WI lamp
has been used. The memory is non-volatile and will stay in the memory even
after power off.
After replacing a new WI lamp or D2 lamp, press <1> or <2> to reset time of
WI lamp or D2 lamp to zero.
3.8.4 GLP/GMP
Press <4> key on the keyboard from the Calibration submenu. Then the time
setting submenu appears on the right of the screen as shown below.