3.8 Calibration
Press [6] on the keyboard to enter the Calibration mode. The Calibration
submenu will be displayed on the LCD as shown below.
11/12/02 10:52 600.0 nm 0.854 ABS
Calibration
1 WL Calibration
2 Set Time
3 Lamp Time
4 GLP/GMP
Select Function: _
3.8.1 Calibrating the Wavelength
The entire calibration procedure is automatic and uses the 656.1 nm emission
lines from the deuterium lamp. Use the following steps to calibrate the
wavelength.
1. Press <1> on the keyboard.
2. Ensure that the cuvette holders are empty and the Sample Compartment is
closed. Press Start.