measurement. The wavelength is selectable with a range of 325.0
~ 370.0 nm in increments of 0.1 nm.
WI Lamp: Turn on or turn off the WI lamp.
D2 Lamp: Turn on or turn off the D2 lamp.
3.6.2 Performing Multiple WL Measurement
Creating a Standard Curve
After setting the appropriate operating conditions, use the following steps to
perform multiple wavelength measurements.
1. Press <0> on the keyboard. The condition-setting screen will switch to the
measurement screen.
2. Press <Autozero> to start an autozero measurement.
3. Insert the first standard sample in the sample compartment. Then press Start
to start measurement.
Repeat the step, as many times as are required to analyze the number of
standards after completing analysis of standard samples, the measurement
screen will change to following screen.
1 Measure 2 Save Inside 3 Save to USB 4 Graph 5 Blank 6 Statistic
16:42 400.0 nm 0.457 ABS
Multiple WL
STD WL1 W
L
Diff Conc
(200nm) (300nm)
1 0.000 0.
0
0.000 0.000
2 0.098 0.10.016 0.005
3 0.203 0.
2
0.017 0.100
Select Function: _
11/13/02
Press <3> key on the keypad to select Graph. Then the new curve will be
created as following.
3-53