11/10/02 17:54 600.0 nm 0.854 ABS
Zoom
0.410
1 X - Max 0.200
2X- Min 0
3 Y- Max 0.410 ABS
4 Y- Min -0.068
-0.068
0.000 MG/L 0.200
Select Function: _
Press <3> key to display the measurement data of standard sample. But it just
displays the mean value of several results for the same standard sample even if it
is measured more than one times. Then select “Remeasure” by pressing <1> on
the keypad to remeasure the absorbance of this standard sample. Select the
“Delete” to delete the result you don’t want.
1 Remeasure 2 Delete
11/12/02 15:41 600.0 nm 0.854 ABS
Photometry
STD ABS Conc
1 -0.068 0.0
20.0981.0
31.0255.0
1-1 -0.068
2-2 -0.068
Select Function: _
If you need measure blank sample value, press <5> before measure sample.
Once you press <5> key, the display will prompt you to insert the sample and
press Start to measure and press Clear/Return to return the last menu. Then
3-17