b) User 1 (calibration according to input conditions)
c) User 2 (calibration according to input conditions)
d) None There is no baseline.
Response:
There are three response speeds to select: Fast, Medium and Slow.
In the ordinary measurement, select “Medium”. If it is desired to reduce
variation, choose the slow response parameter. If it is required carry out high-
resolution measurement, select “Fast” response parameter.
Init Delay:
Specify an initial delay time where it is required to wait until the temperature
becomes stable for sample measurement or the sample sipper is employed.
In an ordinary measurement application, set an initial delay time to ‘0’ (zero).
This parameter is specifiable within a range of 0 to 9999 seconds in increments
of one second.
5. Save Parameter
Press <5> to select Save Parameter from the wavelength scan submenu, a submenu
list appears on the right side of the screen as shown in the illustration below:
10/11/02 13:28 600.0 nm 0.854 ABS
Wavelength Scan
1 Parameter Setup
2 Sample Setup
3 Print Setup
4 System Setup
5Save Parameter
6 User Baseline
0 Forward
Input File No.:_
(1 to 20)
Use the keypad to select the desired catalog number (1-20), then press ENTER. If
the memory location that you have selected is available, the instrument stores entire
variable set under test name you have specified. If the catalog number you
requested is already full, the following message appears:
Overwrite this File?
3-26