EasyManuals Logo

Eaton SEFELEC 5 Series User Manual

Eaton SEFELEC 5 Series
209 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
User manual SEFELEC 5x Series 98 v1.06
THE FOLLOWING ERROR MESSAGES:
"SAFETY LOOP OPEN": the safety loop is not closed, there is no connection between
the points 1-9 or 2-10 of connection C5 on the rear panel of the device. Press OK to
acknowledge the message
VOLTAGE ERROR": The measurement voltage has not reached its final value. The
voltage generator charges the capacity of the sample under test. This message should
appear momentarily when switching to measuring mode. If this message persists the
measurement voltage requested cannot be supplied to the impedance measured.
Check in relation to the characteristics of the voltage generator (difference of 20% in rela-
tion to the set point)
"OVER HEATING": The device is equipped with a thermal protection system. In case of
extended use of the High-voltage generator on a short-circuit, the protection system will
be activated.
In this case the message will be displayed permanently on the LCD screen.
Switch off the device and let it cool down for 10 minutes prior to resuming use.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eaton SEFELEC 5 Series and is the answer not in the manual?

Eaton SEFELEC 5 Series Specifications

General IconGeneral
BrandEaton
ModelSEFELEC 5 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals