Calibration Test Reports
Ectron Model 1140A Calibration Test Report
Thirty-day Specifications, ITS-90
Customer Date
Address Serial Number
Report Number
Other Identifier
Procedure Model 1140A Instruction Manual, Section XI
Linear-voltage Tests
Expanded
Source Meter Measurement
Voltage Reading Reading Tolerance Uncertainty
11.00000 V V V 10.999779 V to 11.000221 V
11.00000 V V V 11.000221 V to 10.999779 V
10.00000 V V V 10.000201 V to 9.999799 V
10.00000 V V V 9.999799 V to 10.000201 V
5.000000 V V V 4.999899 V to 5.000101 V
5.000000 V V V 5.000101 V to 4.999899 V
2.500000 V V V 2.500051 V to 2.499949 V
2.500000 V V V 2.499949 V to 2.500051 V
1.377000 V V V 1.3769715 V to 1.3770285 V
1.377000 V V V 1.3770285 V to 1.3769715 V
1.373000 V V V 1.3730285 V to 1.3729715 V
1.373000 V V V 1.3729715 V to 1.3730285 V
500.0000 mV mV mV 499.9890 mV to 500.0110 mV
500.0000 mV mV mV 500.0110 mV to 499.9890 mV
88.0000 mV mV mV 88.0028 mV to 87.9972 mV
88.0000 mV mV mV 87.9972 mV to 88.0028 mV
85.0000 mV mV mV 84.9973 mV to 85.0027 mV
85.0000 mV mV mV 85.0027 mV to 84.9973 mV
0.0000 mV µ V µ V 1.0 µV to 1.0 µV