Specifications
Table 2-17: Type Platinel II Thermocouple (°C)
Temperature Range
Error After:
30 Days Six Months One Year
0 <100
±0.07 ±0.08 ±0.10
100 <925
±0.06 ±0.07 ±0.08
925 <1200
±0.07 ±0.08 ±0.10
1200 1395
±0.08 ±0.09 ±0.11
Table 2-18: Type Platinel II Thermocouple (°F)
Temperature Range
Error After:
30 Days Six Months One Year
32 <212
±0.13 ±0.14 ±0.18
212 <1697
±0.11 ±0.13 ±0.14
1697 <2192
±0.13 ±0.14 ±0.18
2192 2543
±0.14 ±0.16 ±0.20
Table 2-19: Type R Thermocouple (°C)
Temperature Range
Error After:
30 Days Six Months One Year
−50 <−30 ±0.40 ±0.58 ±0.65
−30
<45
±0.34 ±0.48 ±0.55
45 <160
±0.24 ±0.32 ±0.40
160 <380
±0.18 ±0.26 ±0.30
380 <775
±0.15 ±0.21 ±0.26
775 1768.1
±0.13 ±0.18 ±0.22
Table 2-20: Type R Thermocouple (°F)
Temperature Range
Error After:
30 Days Six Months One Year
−58 <−22 ±0.72 ±1.04 ±1.17
−22
<113
±0.61 ±0.86 ±0.99
113 <320
±0.43 ±0.58 ±0.72
320 <716
±0.32 ±0.47 ±0.54
716 <1427
±0.27 ±0.38 ±0.47
1427 3214.58
±0.23 ±0.32 ±0.40
2-8