EasyManua.ls Logo

Ectron 1140A - Table 2-21: Type S Thermocouple (°C); Table 2-22: Type S Thermocouple (°F); Table 2-23: Type T Thermocouple (°C); Table 2-24: Type T Thermocouple (°F)

Ectron 1140A
136 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
Table 2-21: Type S Thermocouple (°C)
Temperature Range
Error After:
30 Days Six Months One Year
−50 <−30 ±0.38 ±0.53 ±0.62
−30
<45
±0.32 ±0.47 ±0.56
45 <105
±0.23 ±0.34 ±0.40
105 <310
±0.20 ±0.30 ±0.33
310 <615
±0.17 ±0.25 ±0.29
615 1768.1
±0.15 ±0.22 ±0.26
Table 2-22: Type S Thermocouple (°F)
Temperature Range
Error After:
30 Days Six Months One Year
−58 <−22 ±0.68 ±0.95 ±1.12
−22
<113
±0.58 ±0.85 ±1.01
113 <221
±0.41 ±0.61 ±0.72
221 <590
±0.36 ±0.54 ±0.59
590 <1139
±0.31 ±0.45 ±0.52
1139 3214.58
±0.27 ±0.40 ±0.47
Table 2-23: Type T Thermocouple (°C)
Temperature Range
Error After:
30 Days Six Months One Year
−270 <−255 ±1.40 ±1.60 ±1.80
−255 <−240 ±0.27 ±0.35 ±0.49
−240 <−210 ±0.17 ±0.24 ±0.30
−210 <−150 ±0.11 ±0.15 ±0.18
−150 <−40 ±0.08 ±0.10 ±0.12
−40
<100
±0.06 ±0.07 ±0.08
100 400
±0.05 ±0.06 ±0.07
Table 2-24: Type T Thermocouple (°F)
Temperature Range
Error After:
30 Days Six Months One Year
−454 <−427 ±2.52 ±2.88 ±3.24
−427 <−400 ±0.49 ±0.63 ±0.88
−400 <−346 ±0.32 ±0.43 ±0.54
−346 <−238 ±0.20 ±0.27 ±0.31
−238 <−40 ±0.14 ±0.18 ±0.20
−40
<212
±0.11 ±0.13 ±0.14
212 752
±0.09 ±0.11 ±0.13
2-9

Table of Contents