EasyManua.ls Logo

Ectron 1140A - Section X, Alignment; General Alignment Process

Ectron 1140A
136 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SECTION X
10 ALIGNMENT
GENERAL
The Model 1140A is a self-aligning instrument. Once VOLTAGE ALIGNMENT and TERMINAL
ALIGNMENT have been performed, the instrument is ready for use. There are no potentiometers to
set or resistors or capacitors to select. Allow a 30 minute warm-up before performing the
alignment procedure.
EQUIPMENT REQUIRED
Ten-volt dc standard: Fluke Model 732B or equivalent. The uncertainty of the voltage
must be <75 µV (for a minimum test-accuracy ratio, TAR, of 4:1) and be within the range
of 9.9 V dc to 10.1 V dc.
Shorting bar or cable: Pomona Electronics Model 5145 or equivalent.
Null detector/microvoltmeter: Keithley Model 155 or equivalent. This unit must be
capable of resolving 1 µV dc.
Low-thermal cable, Pomona 1756-24 or equivalent.
Ice-point bath: Hart Model 9101 zero-point dry well or equivalent. A properly made and
maintained distilled-water ice bath can also be used. An uncertainty of 0.01°C is required.
Calibrated Type T thermocouple (24 AWG recommended) with its error at 26°C
3
, if any,
known to the nearest tenth of a microvolt. Appendix B provides a procedure to calibrate a
thermocouple. If copper wires are attached to the thermocouple, the copper ends should
be shorted together. The insulation on the thermocouple wire ends should be stripped
back approximately 0.4 (10 mm).
Optional: Cover to shield terminals from air flow, such as Ectron P/N 114-523-01.
For units with firmware version 4.40 and above, the following items from the Calibration
Procedure are also needed during alignment:
Digital multimeter (DMM), Hewlett Packard Model 3458A (with Option 02) or
equivalent.
Four-wire Type E thermocouple (24 AWG recommended), calibrated at 26°C
3
with
known microvolt error and with bare-wire ends for alloy and copper wires. Appendix
B provides a procedure to calibrate a thermocouple. The insulation on the
thermocouple wire ends should be stripped back approximately 0.4 (10 mm).
PRELIMINARY
Turn the power to the Model 1140A off, set the alignment switch (recessed in a square hole in the
bottom cover just to the rear of the KEYPAD) to the right to enable ALIGNMENT, and turn the
Model 1140A on.
3 See Footnote 5 in Appendix B.
10-1

Table of Contents