EasyManua.ls Logo

Ectron 1140A - Table 2-13: Type K Thermocouple (°C); Table 2-14: Type K Thermocouple (°F); Table 2-15: Type N Thermocouple (°C); Table 2-16: Type N Thermocouple (°F)

Ectron 1140A
136 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
Table 2-13: Type K Thermocouple (°C)
Temperature Range
Error After:
30 Days Six Months One Year
−270 <−255 ±1.50 ±1.90 ±2.20
−255 <−195 ±0.30 ±0.40 ±0.70
−195 <−115 ±0.10 ±0.11 ±0.12
−115 <−55 ±0.07 ±0.08 ±0.09
−55
<1000
±0.06 ±0.07 ±0.07
1000 1372
±0.06 ±0.08 ±0.08
Table 2-14: Type K Thermocouple (°F)
Temperature Range
Error After:
30 Days Six Months One Year
−454 <−427 ±2.70 ±3.42 ±3.96
−427 <−319 ±0.54 ±0.72 ±1.26
−319 <−175 ±0.18 ±0.20 ±0.22
−175 <−67 ±0.13 ±0.14 ±0.16
−67
<1832
±0.11 ±0.13 ±0.13
1832 2501.6
±0.11 ±0.14 ±0.14
Table 2-15: Type N Thermocouple (°C)
Temperature Range
Error After:
30 Days Six Months One Year
−270 <−260 ±3.50 ±4.00 ±5.00
−260 <−200 ±0.75 ±0.93 ±1.00
−200 <−140 ±0.15 ±0.19 ±0.23
−140 <−70 ±0.10 ±0.12 ±0.15
−70
<25
±0.08 ±0.10 ±0.12
25 <160
±0.07 ±0.09 ±0.10
160 1300
±0.07 ±0.08 ±0.09
Table 2-16: Type N Thermocouple (°F)
Temperature Range
Error After:
30 Days Six Months One Year
−454 <−436 ±6.30 ±7.20 ±9.00
−436 <−328 ±1.35 ±1.67 ±1.80
−328 <−220 ±0.27 ±0.34 ±0.41
−220 <−94 ±0.18 ±0.22 ±0.27
−94
<77
±0.14 ±0.18 ±0.22
77 <320
±0.13 ±0.16 ±0.18
320 2372
±0.13 ±0.14 ±0.16
2-7