Patient Monitor User Manual Monitoring C.O.
- 98 -
C.O. Measure Window
Measurement curve
Prompt message area
Cardiac Output
Cardiac Index
Body Surface Area
Blood Temperature
Injectate Temperature
Start time of the measurement
Funciton keys
The functional keys on the C.O. measure window are explained in the following table:
Start
Start a measurement
Stop
If the blood temperature cannot resume in a considerably
long time, the measurement could not stop automatically.
Use this button to stop the measurement and display the
C.O., CI calculation result.
Cancel
Cancel the processing measurement or cancel the result
after measurement.
Record
Print out the curve.
Y axis
Change the scale Y (temperature) value. Three modes are
available: 0~0.5°C, 0~1°C, 0~2.0°C. Adjust the scale by
the temperature differences. A
smaller scale results in a
larger curve.