Patient Monitor User Manual Maintenance
- 135 -
NIBP Leakage Inspection
At least once every two years, or as specified by local
laws.
NIBP Pressure Calibration
At least once every two years, or as specified by local
laws.
NIBP Calibration
At least once every two years, or as specified by local
laws.
CO
2
Calibration and Performance
Test
At least once every two years, or if you suspect the
measurement values are incorrect.
AG Calibration
If you suspect the measurement values are incorrect and
need to calibrate, please contact the manufacturer.