Patient Monitor User Manual Product Specification
- 159 -
± 5 % of reading, 41 to 70 mmHg
± 8 % of reading, 71 to 100 mmHg
± 10 % of reading, 101 to 150 mmHg
10s, 15s, 20s, 25s, 30s, 35s, 40s, default value is 20
Drift over 4 hours < 0.8 mmHg
It displays the value within 15s and meets the requirement
for measurement accuracy within 2min. (Mainstream)
It displays the value within 20s and meets the requirement
for measurement accuracy within 2min. (Sidestream)
10s, 15s, 20s, 25s, 30s, 35s, 40s, 45s; default value is 20s.
Interfering Gas and Vapor Effect on EtCO
2
Measurement Values:
Quantitative effect/Comments
0 – 40 mmHg: ± 1 mmHg additional error