Patient Monitor User Manual Product Specification
- 165 -
CO
2
±(0.3 vol% + 4 % of reading)
N
2
O
±(2 vol% + 4 % of reading)
Agents
±(0.2 vol% + 10 % of reading)
Apnea Alarm
Delay
20s ~ 60s
Temperature(°C)
10 to 40
-20 ~ 75
Relative
humidity
10 ~ 95% RH(non-condensing )
5 ~ 100% RH(condensing )
Atmospheric
pressure
525 ~ 1200hPa
500 ~ 1200hPa
Interfering gas and vapour effects
Gas or vapour
Gas level
CO
2
Agents
N
2
O
IRMA CO
2
IRMA AX+
N
2
O
4)
60 vol%
_ 1&2)
_ 1&2)
_ 1)
_ 1)
HAL
4)
4 vol%
_ 1)
_ 1)
_ 1)
_ 1)
ENF, ISO, SEV
4)
5 vol%
+8% of
reading
3)
_ 1)
_ 1)
_ 1)
DES
4)
15 vol%
+12% of
reading
3)
_ 1)
_ 1)
_ 1)
Xe(Xenon)
4)
80 vol%
-10% of
reading
3)
_ 1)
_ 1)
He(Helium)
4)
50 vol%
-6% of
reading
3)
_ 1)
_ 1)
Metered dose
inhaler
propellants
4)
Not for use with metered dose inhaler propellants
C
2
H
5
OH(Ethanol)
4)
0.3 vol%
_ 1)
_ 1)
_ 1)
_ 1)
C
3
H
7
OH
(Isopropanol)
4)
0.5 vol%
_ 1)
_ 1)
_ 1)
_ 1)
CH
3
COCH
3
(Acetone)
4)
1 vol%
_ 1)
_ 1)
_ 1)
_ 1)
CH
4
(Methane)
4)
3 vol%
_ 1)
_ 1)
_ 1)
_ 1)
CO(Carbon
monoxide)
5)
1 vol%
_ 1)
_ 1)
_ 1)
_ 1)