Patient Monitor User Manual Alarm Information
- 34 -
NIBP Pneumatic Leak
NIBP cuff or pump has a
leakage.
Low
Check the NIBP cuff
and pum
p for
leakages.
NIBP System Failure Hardware abnormal High
Contact your service
personnel.
NIBP Cuff Leak
Cuff, pump or airway has a
leakage.
Low
NIBP cuff isn’t
properly connected,
or there is a leak in
the airway.
NIBP Leak Test Error Hardware abnormal High
Check whether the
airway is occluded or
pressure sensor
works properly in
pressure meter mode.
If the problem still
exists, contact your
service personnel.
NIBP Weak Signal
Cuff is too loose or patient
pulse is too weak.
Low
Use other method to
m
easure blood
pressure.
NIBP Excessive
Motion
Due to arm motion, signal
noise is too large or pulse
rate is not regular.
Low
Make sure that the
patient under
monitoring is
motionless.
NIBP Range Exceeded
Maybe the patient blood
pressure value is beyond the
measurement range.
Low
Maybe the patient
blood pressure value
is beyond the
measurement range.
NIBP Air Leak
The cuff pressure can’t reach
the set value within 60sec
and 20sec in Neo mode.
Low
Check the
connections and the
wrapped cuff to see
whether they are all
prepared well.
NIBP Pressure Low
The module isn’t
able to
detect the SYSTOLIC.
Apply pressure again.
Low
Inflate again and
retry thrice. Check
whether the patient
has an
over high
blood pressure or it
is interfered by
movement.