Patient Monitor User Manual Alarm Information
- 44 -
IBP alarm limits are listed as follows: unit (mmHg)
ALM HI
ALM LO
Art
300
0
RAP
40
-10
LAP
40
-10
CVP
40
-10
PA
120
-6
ICP
40
-10
P1
300
-50
P2
300
-50
CO
2
alarm limits are listed as follows:
ALM HI
ALM LO
EtCO
2
150 mmHg
0
FiCO
2
50 mmHg
3
AwRR
150 rpm
2 rpm
C.O. alarm limits are listed as follows:
ALM HI
ALM LO
TB
43°C (109.4) 23°C(73.4)
Quick Temp alarm limits are listed as follows:
Patient Type
ALM HI
ALM LO
ADU
42°C (107.6
)
35.5°C (95.9
)
PED
42°C (107.6) 35.5°C (95.9)
AG alarm limits are listed as follows:
Patient Type
ALM HI
ALM LO
ADU
FiCO
2
EtCO
2
FiO
2
EtO
2
FiN
2
O
EtN
2
O
EtDes
FiDes
25.0%
25.0%
88.0%
90.0%
100.0%
100.0%
18.0%
18.0%
0.0%
0.0%
18.0%
18.0%
0.0%
0.0%
0%
0%