Purge gas specic
aon Reference data
Maximum concentraon of oil 0.1 parts per million
Permied purge gas ow (when necessary)
20 to 50 sccm (0.33 to 0.84 mbar l s
-1
 or
33 t
o 84 Pa l s
-1
)
Recommended purge gas o
w
25 sccm (0.42 mbar l s
-1
, 42 P
a l s
-1
)
Maximum permied supply pressure of
the purge gas
1 bar (gauge), 14.5 psig, 2 x 10
5
 P
a
3.5  Cooling water
R
efer to Table: Cooling water specicaon on page 24 for the cooling water
specicaon. The cooling water supply must be a high‑quality drinking water
specicaon. Check with the water supply authority if there is doubt about the quality
of the supply.
Table 7
Cooling water specicaon
Cooling water specicaon Reference data
Quality
Mechanically clean and opcally clear
with no deposits or turbidity
pH value 6.0 to 8.0
Maximum calcium carbonate
concentraon
75 parts per million
Maximum chloride concentraon 100 parts per million
Minimum oxygen concentraon 4 parts per million
Minimum cooling water ow rate (at
15 °C)
15 l hr
-1
Water temperature 10 to 20 °C
Maximum water pressure
5 bar (gauge), 73.5 psig, 6 x 10
5
 P
a
Materials exposed to the cooling water Nickel plated brass
3.6  Electrical data
The pump
s can be operated by the customers system or by the manufacturer's TIC Turbo
Instrument Controller or TIC Turbo Controller.
If using the customer system, the size of the power supply depends on the applicaon.
The rate of increase in the pump speed is dependent on the power limit seng. The
power limit seng overrides the necessary power supply seng. If the serial
communicaons or an access to the manufacturer's TIC is available, the power limit
seng of the  pump can be selected. Refer to Table: Logic interface technical data on
page 27 for the maximum power limit sengs for the pumps. If the rapid cycling of the
pump is necessary for an applicaon, you can get the faster rate of increase in the pump
speed if the power supply supplies higher current, up to a maximum in accordance with
Table: Logic interface technical data on page 27.
Page 24
B85200880_D - Technical data