To check what normal speed is set, send a query as follows:
Command ? S 8 5 6 cr
The reply is as follows:
Reply = S 8 5 6 sp d d d cr
5.1.6  Timer seng and opons
R
efer to Timer on page 12 for a full descripon of mer funconality.
The meout period is a user‑congurable opon and can be set to a value from
1 to 30 minutes. When the pump is shipped, it is congured with a default meout
period of 8 minutes.
To change the mer seng, send the command that follows (where the characters 'd'
represent the meout period in minutes):
Command ! S 8 5 4 sp d d cr
The reply will be as follows:
Reply * S 8 5 4 sp r cr
The mer seng is now stored in memory of the pump.
To check what meout period is set, send a query as follows:
Command ? S 8 5 4 cr
The reply will be as follows:
Reply = S 8 5 4 sp d d cr
The mer is permanen
tly enabled during ramp‑up. It is oponal to have the mer
enabled at other mes. When the pump is shipped, the mer is enabled by default.
To disable the mer, send the serial command that follows:
Command ! S 8 7 0 sp 0 cr
The reply will be as follows:
Reply * S 8 7 0 sp r cr
The state of the mer opon is stored in memory of the pump.
To enable the mer again, send the serial command that follows:
Command ! S 8 7 0 sp 1 cr
The reply will be as follows:
Reply * S 8 7 0 sp r cr
To check if the mer is enabled or disabled, send the quer
y that follows:
Command ? S 8 7 0 cr
The reply will be as follows (where d=0 means disabled and d=1 means enabled):
Page 60
B85200880_D - Con
guraon