Edwards nEXT200200 - Pump Operation Modes

Edwards nEXT200200
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
To check what normal speed is set, send a query as follows:
Command ? S 8 5 6 cr
The reply is as follows:
Reply = S 8 5 6 sp d d d cr
5.1.6 Timer seng and opons
R
efer to Timer on page 12 for a full descripon of mer funconality.
The meout period is a usercongurable opon and can be set to a value from
1 to 30 minutes. When the pump is shipped, it is congured with a default meout
period of 8 minutes.
To change the mer seng, send the command that follows (where the characters 'd'
represent the meout period in minutes):
Command ! S 8 5 4 sp d d cr
The reply will be as follows:
Reply * S 8 5 4 sp r cr
The mer seng is now stored in memory of the pump.
To check what meout period is set, send a query as follows:
Command ? S 8 5 4 cr
The reply will be as follows:
Reply = S 8 5 4 sp d d cr
The mer is permanen
tly enabled during rampup. It is oponal to have the mer
enabled at other mes. When the pump is shipped, the mer is enabled by default.
To disable the mer, send the serial command that follows:
Command ! S 8 7 0 sp 0 cr
The reply will be as follows:
Reply * S 8 7 0 sp r cr
The state of the mer opon is stored in memory of the pump.
To enable the mer again, send the serial command that follows:
Command ! S 8 7 0 sp 1 cr
The reply will be as follows:
Reply * S 8 7 0 sp r cr
To check if the mer is enabled or disabled, send the quer
y that follows:
Command ? S 8 7 0 cr
The reply will be as follows (where d=0 means disabled and d=1 means enabled):
Page 60
B85200880_D - Con
guraon