Error ash
posion
Error ash
c
ode
Comments Acons
2 sLssss
Controller failed internal
conguraon and calibraon
operaon
Restart the system. If the
error code is not cleared,
contact us or the supplier.
3 ssLsss
Failure to reach or maintain
half full speed in the mer
seng value
Check if the pump is too hot
or if the inlet pressure is too
high.
4 sssLss
Overspeed or overcurrent trip
is acvated, or other
hardware fault
Restart the system. If the
error code is not cleared,
contact us or the supplier.
5 ssssLs
Pump internal measurement
s
ystem is disconnected or
damaged
Restart the system. If the
error code is not cleared,
contact us or the supplier.
6 sssssL
Serial enable becomes
inacve aer a serial start
command
Acvate the Serial Enable
again and send a Serial Stop
command to clear the error
code.
Note:
The alarm LED error ash sequenc
e can signal mulple fail condions. For example, error
ash code sLssLs signies error 2 (controller failed internal conguraon and calibraon
operaon) and error 5 (pump internal measurement system disconnected or damaged).
8.9  Decoding system status words
When you use the serial communicaons link, addional informaon that can be useful for the fault
nding can accessed. When sending a query to monitor measured motor speed, the pump also returns
a System Status Word.
The send command that follows:
Command ? V 8 5 2 cr
The reply is as follows (where, r
st returned number refers to motor rotaonal speed in
revoluons per second (Hz)):
Reply = V 8 5 2 sp d d d d ; h h h h h h h h cr
The system status word returned is made up of 8 hexadecimal digits. To decode this
word, convert each digit into a 4‑digit binary number. Example:
2 2 8 3 0 0 2 2
ê ê ê ê ê ê ê ê
0 0 1 0 0 0 1 0 1 0 0 0 0 0 1 1 0 0 0 0 0 0 0 0 0 0 1 0 0 0 1 0
Refer to T
able: Hexadecimal conversion table on page 83 for more informaon.
Table 26
Hexadecimal conversion table
Hexadecimal Binary Decimal
0 0000 0
Page 83
B85200880_D - Fault nding