Edwards nEXT200200 - Controller Run Time; Pump Run Time; Pump Cycles

Edwards nEXT200200
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Bit number
Status ag Acve ag means
7 Service due
Service is due. Specic oper
aon necessary
should be determined by checking the bits
above
8 - 31 Reserved -
8.3 Controller run me
The run hour
s and recommended service me of the controller on the pump can be
monitored.
Send the query that follows:
Command ? V 8 8 1 cr
The reply is as follows (where, rst number is the hours run by the controller and second
is the number of hours unl the service is recommended):
Reply = V 8 8 2 sp d d d d d d ; d d d d d d cr
Note:
The number of hours unl the ne
xt service is due is esmated by the controller based on
the operang condions of the pump and they can decrease at more or less than 1 hour
per hour. Decreased pump temperature is the primary factor in extending the controller
life.
8.4 Pump run me
The run hours and recommended service me of the impeller in the pump can be
monitored.
Send the query that follows:
Command ? V 8 8 3 cr
The reply will be as follows (where, r
st number is the hours run by the impeller and
second is the number of hours unl service is recommended):
Reply = V 8 8 3 sp d d d d d d ; d d d d d d cr
8.5 Pump cycles
The number of s
tart-stop cycles completed and the number remaining unl the next
service is due can be monitored.
Send the query that follows:
Command ? V 8 8 4 cr
The reply is as follows (where, r
st number is the start-stop cycles completed by the
pump and second is the number of start-stop cycles unl service is recommended):
Reply = V 8 8 4 sp d d d d d d ; d d d d d d cr
Page 81
B85200880_D - Fault nding