Remedy Clean the high vacuum system.
Cause Check the system for leaks and c
ontaminaon.
Remedy If a leak is found, repair it and clean the c
ontaminaon.
Cause The inlet pressure is less.
Remedy
• Remove the pump from the system and test the ulma
te pressure of the pump
only.
• Check the pump for contaminaon.
Do the leak test of the pump. If the leak rate more than
1x10
-7
mbar l s
-1
(1x10
-5
Pa l s
-1
), contact us or the supplier.
Fault The pump is very noisy or there is excessive vibr
aon or both
Cause The r
otaonal speed of the pump is same as the resonant frequency of the aached
system.
Remedy Change the natural frequency of the system or isolate the pump using exible bellows.
Cause The vibr
aon is transming from the rotary backing pump.
Remedy Install a exible bellows or a vibraon isolator in the backing line.
Cause The noise is irregular and increasing.
Remedy A de
fecve bearing. Contact us or the supplier.
Cause The pump makes a constant high pitched noise.
Remedy The rotor is out of balance. Contact us or the supplier.
Fault No serial comms
Cause No electrical supply, loss of serial link.
Remedy Check that:
• the electrical supply is on.
• the serial link is connected.
• the serial enable line is acv
e.
Cause Slide switch is not in the correct posion f
or RS232 or RS485.
Remedy Check that:
• the electrical supply is on.
• the serial link is connected.
• the serial enable line is acve.
• Le for RS485, right for RS232.
• Right for parallel control and monitoring.
Cause Incorrect baud rate and node address.
Remedy Check baud rate and, if oper
ang in mul-drop mode, the node address matches those
of the pump.
Cause Check that the red light is not on or ashing.
Remedy If LED is on, check the red alarm LED symptoms.
Page 78
B85200880_D - Fault nding