Object name Command
Parameter
r
ange
Factory
seng
Data type Units
Comments
Standby
speed seng
!S857
?S857
55.100 70 decimal % Standby speed as a percentage
of full speed
Temperature
readings
?V859 0.100
0.100
- decimal
decimal
°C
°C
Measured motor temperature
Measured controller
temperature
Link
parameter
readings
?V860 0.500
0.300
0.15000
- decimal
decimal
decimal
0.1 Volts
0.1 Amps
0.1 Was
Measured link voltage
Measured link current
Measured link power
Factory
sengs
!S867 1 - - - Reset all c
onguraon opons
and parameters to the factory
sengs
PIC soware
version
?S868 10 - string chars Boot loader soware version
number (D374796XXX)
Speed
control
!C869 0
1
- decimal - Set target speed to full speed
Set target speed to standby
speed
Timer
opons
!S870
?S870
0
1
1 decimal - Timer = disabled
Timer = enabled
Note that the mer is
permanently enabled on
rampup.
Analogue
signal
opons
!S871
?S871
0
1
2
3
0 decimal - Analogue output = measured
speed
Analogue output = measured
power
Analogue output = measured
motor temp.
Analogue output = measured
control temp.
Close vent
valve
!C875 1 - decimal - Closes the vent valve for delayed
start and overrides the current
vent opon. There is no open
vent valve command but the
stop command (!C852 0) will
clear the override.
Service
status
?V881 32 bits - hex ags Service status word
Controller
run me
?V882 0..999999 - decimal hours Hours run by controller
0..999999 decimal hours Hours unl controller service
due
Pump run
me
?V883 0..999999 - decimal hours Hours run by pump
0..999999 decimal hours Hours unl pump service due
Page 48
B85200880_D - Installaon