When planning to pump hazardous substances with this pump, read the related chapters
in the Sa
fety Booklet and in these Operang Instrucons rst.
Further details can be obtained by contacng us.
2.3 Descripon
The nXDS pump is shown in Figure: nXDS scroll pump on page 11.
The nXDS pump is a dry vacuum pump as all the bearings, with their hydrocarbon
lubricant, are isolated from the vacuum space. The pump is suitable for use on vapour
handling processes and may be used for some pumping applicaons involving corrosive
substances. For informaon on pumping ammable gases, contact us.
The body of the pump includes a xed scroll and an orbing scroll. The orbing scroll is
controlled by an electric motor through an eccentric cam on the motor drive sha. The
movement of the orbing scroll, meshed with the xed scroll, forms successive crescent
shaped volumes in the pump. Gas that enters the pump through the inlet is compressed
by the movement of the orbing scroll and swept towards the centre of the xed scroll.
The compressed gas enters the exhaust port near the centre of the xed scroll and is
exhausted from the pump through the outlet.
Refer to Operang and storage condions on page 15 for details of operang condions.
2.4 Pump controller
The integral pump controller manages the supply of current to a three‑phase electric
motor in accordance with operang condions. The controller monitors power and
temperature and will protect the pump in the event of operaon under sustained high
load or under fault condions.
The controller provides the user interface (refer to Figure: nXDS scroll pump on page 11).
The pump may be operated in these modes:
▪ Manually, using the buons on the in
terface panel. Refer to Figure: Quick start
guide (manual control mode) on page 12 and Figure: User interface panel on page
32
▪ Remotely via serial communicaons or digital and analogue process control
(parallel), via the 15‑way D‑type logic interface connector. Refer to Logic interface
on page 11.
Page 10
A73501880_G - Introducon