▪ Ensure the maintenance is done by a suitably trained and supervised technician.
Obe
y local and naonal safety requirements.
▪ Ensure the maintenance technician is familiar with the safety procedures which
relate to the products processed by the pumping system.
▪ Check that all the required parts are available and are of the correct type before
starng work.
▪ Isolate the pump and other components from the electrical supply to prevent
accidental operaon.
▪ Allow the pump to cool for at least 3 hours before starng maintenance work.
6.2 Maintenance plan
More frequent maintenance may be required if the pump is used to pump aggressive
gases or vapours, such as solvents, organic substances and acids, or if the pump is
operated connuously at the higher end of its operang temperature.
Table 14
Maintenance plan
Operaon
Frequency
(months)
Service
indicator
Secon reference
Inspect and clean the inlet
strainer
12 No
Inspect and clean the inlet
strainer on page 39
Inspect and clean the external
fan cover if required
12 No
Clean the external fan cover on
page 39
Check the pump performance 30 Yes
Check the pump performance
(service indicator) on page 40
Replace the pump bearings 60 Yes Replace the p-seals on page 40
Replace the pump controller 120 Yes Replace the p-seals on page 40
Electrical safety check 60 No
Replace the pump controller
(service indicator) on page 41
For service indicator codes, refer to Service indicator codes on page 41.
6.3 Inspect and clean the inlet strainer
Whenever the pump is disconnected from the vacuum system, or on an annual basis, we
recommend:
▪ Removing the inlet strainer from the pump inlet (refer to Figure: nXDS scroll pump
on page 11) and remove any debris that may have accumulated.
▪ Inspecng the inlet strainer and if necessary, clean it with a cleaning soluon
suitable for the substances pumped. Ret the inlet strainer before reconnecng
the pump to the vacuum system. Refer to Connect to the vacuum system on page
28.
6.4 Clean the external fan cover
If the fan cover is not kept clean, the air ow over the pump can be restricted and the
pump may overheat.
Page 39
A73501880_G - Maintenance