4. Ins
tallaon
4.1 Safety
WARNING:
Obe
y the safety instrucons in this secon and take note of appropriate precauons.
If not, injury to people and damage to equipment can result.
Prevent any part of the human body coming into contact with the vacuum.
The nXDS pump is not intended for pumping explosive gases connuously (refer to
ATEX direcve implicaon on page 9).
Ensure that the pump is suitable for the applic
aon. If in doubt, refer our guidelines on
vacuum pump and vacuum system safety, or contact us for advice.
A suitably trained and supervised technician must perform the installaon of the pump.
Obey the safety instrucons listed below, especially when connecng the pump into an
exisng system. Details of the specic safety precauons are given at the appropriate
point in the instrucons.
▪ Wear the appropriate safety clothing if contact with contaminated components is
ancipated. Dismantle and clean contaminated components inside a fume
cupboard.
▪ Vent and purge the vacuum system before starng installaon work.
▪ Ensure that the installaon technician is familiar with the safety procedures that
relate to the products handled by the pumping system.
▪ Disconnect the other components in the pumping system from the electrical
supply to prevent accidental operaon.
4.2 System design consideraons
Consider the following points when designing the pumping system:
We recommend you to use a foreline vacuum isolaon valve to allow the pump to warm
up before pumping condensible vapours or if a vacuum needs to be maintained when
the pump is not running.
Use a suitable valve to isolate the pump from the vacuum system if the pump needs to
warm up before pumping condensable vapours or if vacuum needs to be maintained
when the pump is switched o.
Avoid high levels of heat input into the pump from the process gases, otherwise the
pump may overheat and cause the thermal protecon system to operate.
Ensure that the exhaust pipeline cannot become blocked. If an exhaust isolaon valve is
installed, ensure that the pump cannot be operated with the valve closed. Refer to
Connect to the vacuum system on page 28.
Provide for a purge of inert gas when the pumping system is shut down, to dilute
dangerous gases to safe concentraons. Contact our Applicaon team for further advice
on diluon requirements if required.
Page 25