5.4.1 Hardware c
onguraon
Using the 15‑way D‑type connector (Figure: nXDS scroll pump on page 11, item 7) apply
the following signal conguraons to enable the Analogue Speed Control source (refer
to Table: Logic interface connector pins on page 22):
Connect the Analogue Speed Enable control input (pin 1) to the 0 V Control Reference
(pin 2).
Connect a suitably calibrated analogue voltage source (0 to + 10 V), for example, (DAC)
to the analogue speed control input (pin 9). Alternavely connect the output of a
potenometer referenced to the pump reference voltage (pin 11) to the analogue speed
control input (pin 9). Refer to Figure: Logic interface connecons - analogue speed
control on page 34. The 0 V rail of the external voltage source must be connected to the
0 V Control Reference (pin 2) of the pump controller.
5.4.2 Operaon
▪ A + 10 V input equates to a mechanical running speed which is equal to: 100% of
the default run speed, that is, 30 Hz.
▪ The minimum running speed provided by the Analogue Speed control source, is
clamped at the minimum Standby Speed Seng, that is, approximately 67% of the
default run speed of 20 Hz.
▪ The maximum running speed provided by the Analogue Speed control source is
clamped b
y the maximum Standby Speed Seng, that is, 100% of the default run
speed of 30 Hz.
5.5 Auto-run
The auto‑run seng congures the pump to start at power‑up without any customer
intervenon. This parameter is customer congurable via serial communicaons or using
the START / STOP buons. Holding down either the START or STOP buon, for more than
eight seconds, will enable or disable the auto‑run seng. The status of the auto‑run
seng is visible via the auto‑run LED.
The pump can be stopped using either manual, parallel or serial control modes whilst in
auto‑run.
5.6 Use of gas ballast control
The gas ballast control can be used to opmise the performance of the scroll pump for
the applicaon. The performance characteriscs of the pump with gas ballast on are
shown in Performance characteriscs on page 16. The posion of the gas ballast control
can be changed when the pump is either o or operang.
5.6.1 Gas ballast control
Use the gas ballast control to introduce air into the nal stage of the pump. Use of gas
ballast will reduce the condensaon of vapours in the pump. The condensates would
contaminate the pump.
There are only two posions, 0 and 1. The gas ballast control knob will rotate 360° in
either direcon at 90° intervals.
Page 35
A73501880_G - Operaon