EasyManua.ls Logo

Eilersen 5024G - Display and steady filters; Appendix E - Setting of MAC Address

Default Icon
39 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
WWW.EILERSEN.COM
Display and steady filters
The filter frequency depends on the weight display reading update rate. Examples are given
in the table below:
Update period (ms)
20
100
200
400
Filter
Filter frequency (Hz) and total settling time (ms/s)
Taps
Damping
7
-60dB
12 Hz
140 ms
2,4 Hz
700 ms
1,2 Hz
1,4 s
0,6 Hz
2,8s
9
-60dB
10 Hz
180 ms
2,0 Hz
900 ms
1,0 Hz
1,8 s
0,5 Hz
3,6 s
9
-80dB
12 Hz
180 ms
2,4 Hz
900 ms
1,2 Hz
1,8 s
0,6 Hz
3,6 s
12
-60dB
8 Hz
240 ms
1,6 Hz
1,2 s
0,8 Hz
2,4 s
0,4 Hz
4,8 s
12
-80dB
10 Hz
240 ms
2,0 Hz
1,2 s
1,0 Hz
2,4 s
0,5 Hz
4,8 s
15
-80dB
8 Hz
300 ms
1,6 Hz
1,5 s
0,8 Hz
3 s
0,4 Hz
6 s
17
-60dB
6 Hz
340 ms
1,2 Hz
1,7 s
0,6 Hz
3,4 s
0,3 Hz
6,8 s
21
-80dB
6 Hz
420 ms
1,2 Hz
2,1 s
0,6 Hz
4,2 s
0,3 Hz
8,4 s
25
-60dB
4 Hz
500 ms
0,8 Hz
2,5 s
0,4 Hz
5 s
0,2 Hz
10 s
32
-80dB
4 Hz
640 ms
0,8 Hz
3,2 s
0,4 Hz
6,4 s
0,2 Hz
12,8 s
50
-60dB
2 Hz
1,0s
0,4 Hz
5 s
0,2 Hz
10 s
0,1 Hz
20 s
64
-80dB
2 Hz
1,28 s
0,4 Hz
6,4 s
0,2 Hz
12,8 s
0,1 Hz
25,6 s
67
-60dB
1,5 Hz
1,34 s
0,3 Hz
6,7 s
0,15 Hz
13,4 s
0,075 Hz
26,8 s
85
-80dB
1,5 Hz
1,70 s
0,3 Hz
8,5 s
0,15 Hz
17 s
0,075 Hz
34 s
100
-60dB
1 Hz
2,0 s
0,2 Hz
10 s
0,1 Hz
20 s
0,05 Hz
40 s
Appendix E Setting of MAC Address
The MAC Address is pre-set to a unique value from the pool assigned by the IEEE to Eilersen
Electric.
Eilersen Electric MAC Address ranges:
00-50-C2-C5-30-00 to 00-50-C2-C5-3f-ff
20-85-93-60-00-00 to 20-85-93-6f-ff-ff
The MAC Address is set to a default value within this range and CAN NOT be set to a random