Connection-specific DNS Suffix:
IP Address:192.168.135.35 (note: the last .35can beup to .44)
Subnet Mask: 255.255.255.0DefaultGateway:
If you get the following:
Ethernet adapterLocal Area Connection 1:
IP Address: 0.0.0.0
then the PC has not yet receivedan IP address from the DHCP server and you should wait
(up to 30 seconds) to receive an IP address before attempting to connect to the meter. If
after 30 seconds the PC has not received an IP address from the meter DHCP server or the
IP addressshown above (from ipconfig) is different from the range of 192.168.135.35
through 192.168.135.44, verify that the PC is configured to receive its IP address
automatically (via DHCP).
To ensure connection to the meter from the PC, at the DOS prompt type:
ping 192.168.135.100 <enter>
If the meter is reachable, then you will see a message like:
Pinging 192.168.135.100 with 32 bytes of data:
Reply from 192.168.135.100: bytes=32 time < 10ms TTL=64
etc.
If the meter is not reachable, then you will see something like:
Pinging 192.168.135.100 with 32 bytes of data:
Request Timed Out etc.
Q3. How do I connect to multiple meters via Ethernet when they are on the same LAN?
A3. Before connecting multiple meters via Ethernet on a LAN, each meter must be
configured with a unique user- specified IP address (following the initial communication
quick start instructions in ). Contact your IT department for valid IP addresses for your LAN
and Gateway address if required. Once a meter's IP address is configured, the meter may
be connected to the intranet LAN and accessed using that IP address.
3410 Series meters connected to an intranet LAN should not have their DHCP servers
enabled.
Q4. How do I connect to multiple meters via Ethernet when they are on the same hub but
not connected to an intranet LAN?
A4. The PC may receive its IP address from an external DHCP server; in this case, one and
only one meter must have its DHCP server enabled. This DHCP server will serve up to 10 IP
addresses to PCs attempting to talk to all meters on the hub.
Troubleshooting comm., mech., and elec.
134 3410 Series Gas Ultrasonic Flow meters