Reference Manual
00809-0100-4840, Rev. BA
July 2008
4-17
Rosemount 3100 Series
f) Press the blue button → to confirm the new profile. (The flashing stops.)
g) If the new profile is correct, press the red button
↵ to save. The display
changes to the next menu option.
h) If the new profile is incorrect, press the blue button
→ to exit to the menu.
The “ProF” re-appears; re-start at step (b) or press
↓ for the next menu.
NOTE:
The next menu option will depend upon the flow profile chosen:
• 3/2 or 5/2: the transmitter will automatically calculate the Power factor and
only requires the K factor to be entered. (see “K-factor for the Flow Law
(The 3102/3105)“ on page 4-19.)
• Manning: the next menu option is “LEUEL @ max”
(see “Maximum Level Entry (The 3102/3105)“ on page 4-20.)
• Parshall, FF, or FP: the transmitter will automatically calculate the
appropriate Power factor and K factor, and will set the 4 mA point at zero
flow and the 20 mA point at maximum flow. (See “Setting the Output
Damping (The 3102/3105)“ on page 4-24.)