Reference Manual
00809-0100-4840, Rev. BA
July 2008
4-17
Rosemount 3100 Series
f) Press the blue button to confirm the new profile. (The flashing stops.)
g) If the new profile is correct, press the red button
to save. The display
changes to the next menu option.
h) If the new profile is incorrect, press the blue button
to exit to the menu.
The “ProF” re-appears; re-start at step (b) or press
for the next menu.
NOTE:
The next menu option will depend upon the flow profile chosen:
3/2 or 5/2: the transmitter will automatically calculate the Power factor and
only requires the K factor to be entered. (see “K-factor for the Flow Law
(The 3102/3105)“ on page 4-19.)
Manning: the next menu option is “LEUEL @ max
(see “Maximum Level Entry (The 3102/3105)“ on page 4-20.)
Parshall, FF, or FP: the transmitter will automatically calculate the
appropriate Power factor and K factor, and will set the 4 mA point at zero
flow and the 20 mA point at maximum flow. (See “Setting the Output
Damping (The 3102/3105)“ on page 4-24.)