219
Reference Manual
00809-0100-4021, Rev GC
Appendix B: Product Certifications
April 2014
Product Certifications
N1 ATEX Type n
Certificate: BAS01ATEX3432X [HART]; Baseefa03ATEX0709X [Fieldbus]
Standards Used: EN 60079-0: 2012, EN 60079-15: 2010
Markings: HART: II 3G Ex nA IIC T5/T6 Gc; T6(–40 °C
T
a

50 °C), T5(–40 °C T
a
75
°C); Fieldbus: II 3G Ex nA IIC T5 Gc; T5(–40 °C
T
a
75 °C);
Special Condition for Safe Use (X):
1. When fitted with the transient terminal options, the equipment is not capable of
withstanding the 500 V electrical strength test as defined in clause 6.5.1 of EN
60079-15: 2010. This must be taken into account during installation.
ND ATEX Dust
Certificate Number: FM12ATEX0065X
Standards Used: EN 60079-0: 2012, EN 60079-31: 2009, EN 60529:1991 +A1:2000
Markings: II 2D Ex tb IIIC T130 °C Db, (–40 °C
T
a
70 °C); IP66
See Table B-2 at the end of the Product Certifications section for process temperatures.
Special Conditions for Safe Use (X):
1. See certificate for ambient temperature range.
2. The non-metallic label may store an electrostatic charge and become a source of
ignition in Group III environments.
3. Guard the LCD display cover against impact energies greater than 4 joules.
4. Consult the manufacturer if dimensional information on the flameproof joints is
necessary.
International
E7 IECEx Flameproof
Certificate: IECEx FMG 12.0022X
Standards Used: IEC 60079-0:2011, IEC 60079-1: 2007-04, IEC 60079-31: 2008
Markings: Ex d IIC T6…T1 Gb, T6(–50 °C
T
a
40 °C), T5…T1(–50 °C T
a
60 °C);
Ex tb IIIC T130 °C Db, (–40 °C
T
a
70 °C); IP66;
See Table B-2 at the end of the Product Certifications section for process temperatures.
Special Conditions for Safe Use (X):
1. See certificate for ambient temperature range.
2. The non-metallic label may store an electrostatic charge and become a source of
ignition in Group III environments.
3. Guard the LCD display cover against impact energies greater than 4 joules.
4. Consult the manufacturer if dimensional information on the flameproof joints is
necessary.