220
Reference Manual
00809-0100-4021, Rev GC
Appendix B: Product Certifications
April 2014
Product Certifications
I7 IECEx Intrinsic Safety
Certificate: IECEx BAS 07.0002X [HART]; IECEx BAS 07.0004X [Fieldbus]
Standards Used: IEC 60079-0: 2011; IEC 60079-11: 2011;
Markings: HART: Ex ia IIC T5/T6 Ga; T6(–60 °C
T
a
50 °C), T5(–60 °C T
a
75 °C);
Fieldbus: Ex ia IIC T5/T6 Ga; T4(–60 °C
T
a
60 °C)
See Table B-3 at the end of the Product Certifications section for entity parameters.
Special Conditions for Safe Use (X):
1. When fitted with the transient terminal options, the apparatus is not capable of
withstanding the 500 V electrical strength test as defined in Clause 6.3.13 of IEC
60079-11: 2011. This must be taken into account during installation.
2. The enclosure may be made from aluminum alloy with a protective polyurethane paint
finish; however, care should be taken to protect it from impact or abrasion when
located in Zone 0.
N7 IECEx Type n
Certificate: IECEx BAS 070003X [HART]; IECEx BAS 07.0005X [Fieldbus]
Standards Used: IEC 60079-0: 2011, IEC 60079-15: 2010
Markings: HART: Ex nA IIC T5/T6 Gc; T6(–40 °C
T
a
50 °C), T5(–40 °C T
a
75 °C);
Fiedlbus: Ex nA IIC T5 Gc; T5(–40 °C
T
a
75 °C);
Brazil
E2 INMETRO Flameproof
Certificate: CEPEL 04.0307X
Standards Used: ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-1:2009, ABNT NBR
IEC 60079-26:2008, ABNT NBR IEC 60529:2009
Markings: Ex d IIC T* Gb; T6(–40 °C
T
a
65 °C), T5(–40 °C T
a
80 °C)
Special Conditions for Safe Use (X):
1. The accessory of cable entries or conduit must be certified as flameproof and needs to
be suitable for use conditions.
2. For ambient temperature above 60 °C, cable wiring must have minimum isolation for
temperature 90 °C, to be in accordance to equipment operation temperature.
3. Where electrical entry is via conduit, the required sealing device must be assembled
immediately close to enclosure.
I2 INMETRO Intrinsic Safety
Certificate: CEPEL 05.0723X
Standards Used: ABNT NBR IEC 60079-0:2008, ABNT NBR IEC 60079-11:2009, ABNT NBR
IEC 60079-26:2008, ABNT NBR IEC 60529:2009
Markings: Ex ia IIC T* Ga; T6(–60 °C
T
a
+50 °C), T5(–60 °C T
a
75 °C),
T4(–60 °C
T
a
60 °C); IP66(Aluminum enclosures), IP66W (Stainless Steel enclosures)
See Table B-3 at the end of the Product Certifications section for entity parameters.
Special Conditions for Safe Use (X):
1. The apparatus enclosure may contain light metals. The apparatus must be installed in
such a manner as to minimize the risk of impact or friction with other metal surfaces.
2. A transient protection device can be fitted as an option, in which the equipment will not
pass the 500 V test.