Gammapilot M FMG60 PROFIBUS PA Commissioning
Endress+Hauser 45
"Value full" (*14) / "Value empty" (*17)
Local display
Value full *14
100%
Local display
Value empty *17
0%
These functions are needed for level measurements only. They are used to specify the level at
which the full or empty calibration is performed.
Value range
optimum value minimum value maximum value
Value full (*14) 100% 60% 100%
Value empty (*17) 0% 0% 40%
"Calibration" (*15)
Local display
Calibration *15
Stop/edit
Start
This function is used to start automatic entry of the selected calibration point
Options:
Stop/edit
This option must be selected if:
The calibration point is not to be entered (e.g. because it has already been entered). The
pulse rate of the calibration point is then displayed in the following function "Full
calibration" (*16) or "Empty calibration" (*18). This value can be changed if required.
The calibration point is to be entered manually. The Gammapilot M changes to the "Full
calibration" (*16) or "Empty calibration" (*18) function.
Start
This option is used to start automatic entry of the calibration point. The Gammapilot M
then changes to the "Avg. pulse rate" (*11) function.