Commissioning Gammapilot M FMG60 PROFIBUS PA
46 Endress+Hauser
"Avg. pulse rate" (*11)
Local display
Avg. pulse rate *11
2548 cps
The average (integrated) pulse rate is displayed in this function (after the selection of "Start" in
the previous function). Initially, this value fluctuates (because of the decay statistics), but
reaches an average value over time due to integration. The longer the averaging is performed,
the lower the fluctuations.
c/s
t
"E"
A0018118
 5 The integrated (average) pulse rate fluctuates greatly at the start. An average value is reached over
time.
When the value is sufficiently stable users can quit the function by pressing . After this, the
Gammapilot M changes to the "Calibration" (*15) function. Here the "Stop/edit" option must
be selected to stop the integration. The value is then automatically transmitted to the "Full
calibration" (*16) or "Empty calibration" (*18) function.
Avg. pulse rate
The maximum integration time (= averaging time) is 1000 s. After this time, the value
is automatically transmitted to the "Full calibration" (*16) or "Empty calibration"
(*18) function.
Integration is not terminated by pressing
in the "Avg. pulse rate" (*11) function.
Integration continues until "Stop/edit" is selected in the "Calibration" (*15) function.
For this reason, there may be a slight deviation between the last average (integrated)
pulse rate displayed and the final "Full calibration" (*16) or "Empty calibration"
(*18).
"Full calibration" (*16) / "Empty calibration" (*18)
Local display
Full calibration *16
33 cps