Commissioning Gammapilot M FMG60 PROFIBUS PA
54 Endress+Hauser
an average value over time due to integration. The longer the averaging is performed, the
lower the fluctuations.
A0018118
When the value is sufficiently stable users can quit the function by pressing . After this, the
Gammapilot M changes to the "Calibration" (*15) function. Here the "Stop/edit" option must
be selected to stop the integration. The value is then automatically transmitted to the
"Density adjustment" (*1B) function.
Density calibration
• The maximum integration time (= averaging time) is 1000 s. After this time, the value
is automatically transmitted to the "Density calibration (*1B) function.
• During the integration, a sample of the measured material must be taken. Its density
must then be determined (e.g. in the laboratory).
• Integration is not terminated by pressing
in the "Avg. pulse rate" (*11) function.
Integration continues until "Stop/edit" is selected in the "Calibration" (*15) function.
For this reason, there may be a slight deviation between the last average (integrated)
pulse rate displayed and the final "Density calibration (*1B).
"Density calibration" (*1B)
Local display
Density calibration *1B
1983 cps
The pulse rate of the calibration point is displayed in this function. Press to confirm the
value and complete the entry of the calibration point. If a calibration has not yet been
performed for the current point, "-1" is displayed. There are two options if this occurs:
• Either return to the "Calibration" (*15) function and restart the calibration
• Or enter a known or calculated pulse rate (manual calibration)