Micropilot FMR10 Electrical connection
Endress+Hauser 21
6.3 Connecting the device
A0028907
17 FMR10 block diagram
1 Micropilot FMR10, 4 to 20 mA
2 Power supply
6.4 Post-connection check
Is the device or cable undamaged (visual inspection)?
Do the mounted cables have adequate strain relief?
Are the cable glands mounted and firmly tightened?
Does the supply voltage match the specifications on the nameplate?
No reverse polarity, is terminal assignment correct?