Micropilot FMR10 Diagnostics and troubleshooting
Endress+Hauser 27
9.3 Diagnostic event in the operating tool
If a diagnostic event is present in the device, the status signal appears in the top left status
area of the operating tool along with the corresponding symbol for the event level in
accordance with NAMUR NE 107:
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
Calling up remedial measures
‣
Navigate to the Diagnostics menu
In the Actual diagnostics parameter the diagnostic event is shown with event text
9.4 List of diagnostic events
Diagnostic
number
Short text Remedy instructions Status signal
[from the
factory]
Diagnostic
behavior
[from the
factory]
Diagnostic of electronic
270 Main electronic
failure
Exchange device F Alarm
271 Main electronic
failure
1. Restart device
2. If failure remains, exchange
device.
F Alarm
272 Main electronic
failure
1. Restart device
2. Check enviroment for strong
EMC fields
3. If failure remains, exchange
device.
F Alarm
283 Memory content 1. Transfer data or reset device
2. Contact service
F Alarm
Diagnostic of configuration
410 Data transfer 1. Check connection
2. Retry data transfer
F Alarm
411 Up-/download active Up-/download active, please wait C Warning
435 Linearization Check linearization table F Alarm
438 Dataset 1. Check data set file
2. Check device configuration
3. Up- and download new
configuration
M Warning
441 Current output 1 1. Check process
2. Check current output settings
S Warning
491 Current output 1
simulation
Deactivate simulation C Warning
585 Simulation distance Deactivate simulation C Warning
586 Record map Recording of mapping
please wait
C Warning
Diagnostic of process
801 Energy too low Increase supply voltage S Warning
825 Operating
temperature
1. Check ambient temperature
2. Check process temperature
S Warning