Micropilot FMR10 Operating menu
Endress+Hauser 55
Linearization type
Navigation  Setup → Advanced setup → Linearization type
Description Linearization types
Meaning of the options:
None:
The level is output in the level unit without being converted (linearized) beforehand.
Table:
The relationship between the measured level L and the output value (volume/flow/
weight) is defined by a linearization table. This table consists of up to 32 value pairs, i.e.
"level - volume" or "level - flow" or "level - weight".
Note:
Please use the function in SmartBlue to create/modify a linearization table.
Selection None
Table
Factory setting None
Level linearized
Navigation  Setup → Advanced setup → Level linearized
Description Currently measured level.
User interface Signed floating-point number