34
Prompt
Language
Contrast of the LCD display
Date
Time of day
Measurement type
Measuring cell
Measuring principle
(difference only)
Unit
Temperature sensor 1
Temperature sensor 2
Temperature unit
Temperature compensation
Compensation temperature
(MTC 1 only; MTC 2 for
difference only)
Air set select
Input contact
(with FCXI card only)
Relay function
NAMUR contacts
1)
Failure relay
Fault contact assignment
Proc. Warnings
Set-up guide
Possible Choices
Deutsch, English, Francais, Italian
Set as desired
Entry of current date
Entry of current time of day
Conductivity, Resistance (M Ohm)
Concentration
CLS 50, CLS 52
2-electrode K = 0, 01
K = 0, 1
K = 1
K = 10
Difference measurement (one-circuit of
difference)
Selection of available units: µS/cm,
mS/cm or mS/m, S/m
Type of temperature sensor
Pt 100 / Pt 1000 / NTC 30K
Type of temperature sensor
Pt 100 / Pt 1000 / NTC 30K
(only 2-circuit)
Celsius (°C),
Fahrenheit (°F),
Kelvin (K)
Manual (MTC)
Automatic (ATC)
-31.0 to +482.0°F
(-35.0 to +250.0°C)
Skip Air Set
Activate Air Set
Input Air Set
2 x hold;
1 x hold + 1 x remote switching; 2 x remote
switching
Fault & Maint.
Fault & Contact
N.C. contact,
N.O. contact
Fleeting contact, Steady contact
Pulsed, continuous
No contact
Assign to maintenance contact
Assign to no contact
End, Resume
Set-up Guide / Checklist
Factory Settings
English
Medium
Current date
CET, (no daylight
savings time)
Conductivity
CLS 52
Difference
measurement
µS/cm, mS/cm
Pt 100
Pt 100
Celsius (°C)
ATC
77°F
(25°C)
Air set escape
2 x hold
N.O. contact
Steady contact
End
User Settings
1)
According to recommendation by NAMUR (Normen-Ausschub Meb - und Regeltechnik, i.e., Measuring and
Control Engineering Standards Committee)