Note!
Measuring cell versions equipped with dairy pipe, clamp and internal thread
fittings can only be installed in pipes with a diameter of at least DN 65.
Evaluation of Installation Factor
Function Selection Default
Measuring cell selection (not for Measuring cell 1 Measuring
one-circuit) Measuring cell 2 cell 1
Note Leave meas.
cell in process
Display of calibration solution, Tc,
temperature
Entry of calibration solution setpoint 0 µS to 1000 mS/cm Current meas.
value
Display of adaptation factor x.xxx
Calibration End cal. End cal.
Repeat cal.
Cal. channel 2 (two-circuit only)
For End cal. selection
Return to measuring mode
For Repeat cal. selection
Calibration of measuring cell 1 Measuring
Return to Measuring cell selection cell 1
(see above)
For Cal. channel 2 selection (two-circuit only)
Calibration of measuring cell 2 Measuring
Return to measuring cell selection cell 2
(see above)
9.2 Contacting Calibration
® Input of cell constant + Chapter 9.2.1
® Evaluation of cell constant + Chapter 9.2.2
9.2.1 Numerical Calibration / Input of Cell Constant
The cell constant is measured exactly at the factory, where it is entered directly in cm
-1
.
The entry is made separately for each measuring cell in the case of two-circuit
measurement.
Cell Constant Entry range
0.01 cm
-1
0.0005 to 0.0500 cm
-1
0.1 cm
-1
0.050 to 0.500 cm
-1
1 cm
-1
0.500 to 5.000 cm
-1
10cm
-1
5.00 to 99.99 cm
-1