55
7.6 Substance Selection / Concentration Measurement
® Commissioning (Start-up)
® Substance selection
Note!
This menu only appears when concentration measurement has been selected as
the measurement type (see Measurement type in chapter 7.1.1).
Function Selection Factory setting
Measuring range selection with Measuring range 1 - 4 Measuring
remote switching activated range 1
Substance selection NaOH NaOH
HNO
3
(fixed, other settings
H
3
PO
4
not required)
H
2
SO
4
USER 1 - USER 4 (user-selectable)
For USER 1 - USER 4
Entry of a substance name Substance designation, max. 5 USER 1
characters: 0 to 9, a to Z
Entry of a number of elements Substance curve elements, 2 - 10 2
(value pairs) for the substance
concentration curve
Each element consists of a Concentration Conductivity
conductivity value and 000.0 % 0.000 mS/cm
associated concentration value
Entry of number of elements Elements of Tc curve, 2 - 10 2
(value pairs) for the Tc curve
Each element consists of a
temperature value and
associated temperature
coefficient
If an error is made entering the numbers, this is indicated by an appropriate error message.